OxiCard® Specifications
/files/Graphic Files/adobe pdf_Resized_17x17.png Printable version of the OxiCard Specifications.

Weight: 170 grams
Dimensions: (W) 43 mm x (D) 36 mm x (H) 3 mm
Housing: CF Card with PCMCIA adapter
Program: Reporting software is stored on the computer
Environmental Conditions: Storage Conditions: Ambient Temperature: -20 to 60° C
Relative Humidity: 5 to 95% (non-condensing)
Atmospheric Pressure: 503 to 1059 hPa
Power Supply: Internal: 3.3Vdc, less than 100mA. Supplied by the PCMCIA slot
Operating Conditions: Ambient Temperature: 0 to 40° C
Relative Humidity: 5 to 95% (non-condensing)
Atmospheric Pressure: 503 to 1059 hPa
Accuracy: SpO2: (± 1 S.D.)* 70-100% ±2% of full scale
Pulse Rate: ±3 beats per minute
Range: % Saturation: 0 to 100%
Pulse Rate: 30 to 235 pulses per minute

*Standard Deviation: (S.D.) is a statistical measure: Up to 32% of the readings may fall outside these readings.